Roll over image to zoom in
Anesthesia Machine NEUGA06
$11,770.00$12,990.00 (-9%)
Main body | High-strength engineering plastic rack, light,
beautiful and corrosion resistant |
|
Scope of application | adult, child | |
Gas source | O2: 0.27~0.55MPa N2O: 0.27~0.55MPa | |
Flow meter |
O2 : 0.1~1.0L/min 1.1~10L/min |
|
N20: 0.1~1.0L/min 1.1~10L/min | ||
O2, N2O linkage and N2O Stopper
When using nitrous oxide, oxygen concentration>25%; When the oxygen pressure<0.2Kpa, the flow of nitrous oxide would be cut off. |
||
Flow rate of rapid oxygen supply | 25~75L/min | |
Low oxygen pressure alarm (Oxygen Alarm) | There will be sound alarm when the oxygen
pressure < 0.2MPa |
|
Vaporizer (Vaporizer Definition and Vaporizer Construction) | it has the function of automatic compensation based on pressure, temperature, and flow rate. The regulation range of evaporator concentration is 0~5 vol%. Among Halothane,Enflurane, Isoflurane and Sevoflurane, two can be chosen for application as required by the customer. And vaporizers that are imported with original packaging are also available. | |
Respiratory circuit | working mode: all-close, semi-close, semi-open
APL: 0.5~7 kPa |
|
Respiratory Bellows | bellows for adults, tidal volume range:0~1500 ml
bellows for children, tidal volume range:0~300 ml anyone can be chosen for application as required by the customer |
|
Ventilator (Ventilator Definition and Ventilator Function) | ||
Display mode | high-definition 5.7” LCD display | |
Ventilation mode | IPPV, SIPPV, SIMV, PEEP, MANUAL, SIGH | |
Ventilation parameters | Tidal volume | 20~1500 ml |
Rate | 1~99 bpm | |
SIMV rate | 2~20 bpm | |
I:E | 4:1~1:8 | |
Inspiratory trigger pressure | -1.0~2.0 kPa | |
PEEP | 0~2.0 kPa | |
Pressure range | 1.0~6.0 kPa | |
SIGH | 1.5 times the inspiratory time
(70-150 adjustable)
|
|
Parameters for ventilation monitoring | tidal volume, ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak pressure of airway, average pressure, pressure – time waveform, flow rate – time waveform,PEEP,inspiratory trigger pressure, Inspiratory platform, autonomous inspiratory rate, lung compliance, airway resistance, flow-volume loop, pressure-volume loop | |
Monitoring of oxygen concentration | (21%~100%) | |
Security Alarm System | ||
Oxygen Concentration alarm | upper limit setting range 21%~100%
low limit setting range 10%~80% |
|
Airway pressure alarm | upper limit setting range 0.3~6.0 kPa
low limit setting range 0.2~5.0 kPa |
|
Per-minute ventilation volume alarm | upper limit setting range 3.0~30L/min
low limit setting range 1.0~10L/min |
|
Sustained high-pressure alarm | it will give alarm when stress have consistently been higher than 2.5 kPa | |
Suffocation alarm | it will give sound and light alarm if there is no tidal volume input for 15 seconds | |
Power alarm | ||
Power | AC 220V±10% 50Hz±1Hz(UPS, Storage battery) |